Signum 3 Ex
Averaging (Animal Weighing) Application V
Averaging (Animal
Weighing) Application V
With the Averaging application, you
can use your weighing platform for
calculating weights as the average
of a number of individual weighing
operations� These individual operations
are also known as “subweighing
operations�
This function is used to determine
weights under unstable ambient
conditions or for weighing unstable
samples (such as live animals)�
Characteristics
Averaging started manually or auto-
matically� Configured in Setup under:
appl1: anim.wg: 3.18.
With manual start selected, the
averaging routine begins when
you press a key (provided the start
conditions are met)� With automatic
start selected, averaging begins when
you place the first load on the platform
(provided the start conditions are met)�
Enter the number of subweighing
operations using the keypad
Info mode
Toggle the display from “result of last
measurement” to current weight” by
pressing the
w key
Autmatic printout configured in Setup
under:
Appl1:
anim.wg: 3.20.
Automatic taring of container weight�
Configured in Setup under:
appl1:
anim.wg
: 3.7.
Automatic start of averaging when
the scale is turned on and a sample
placed on the platform (provided start
conditions are met)� Configured in
Setup under:
Appl: a.start: 3.8.
Exit application, delete parameters:
The number of measurements remain in
memory until it is deleted, overwritten
or the application is changed using the
c key
The number of measurements also
remains saved after the scale is turned
o�
You can assign dierent functions to
the c key for deleting applications�
When you clear applications, you can
delete either the data stored for all
applications or just selected data stored
for the active application� Congured in
Setup under:
Appl: sel.cf: 3.24.
(Sel� CF function key c in
applications)
Tare function:
If you store a tare (weight value) by
pressing the ) key, you can later
enter a value manuallyThe tare value
you enter is added to the stored tare
value�
Setting: menu code 3�25�1 (factory
setting)
A tare value entered manually
overwrites a stored tare value (weight
value)� If you enter a tare value
manually, a tare value (weight value)
stored later overwrites the manually
entered value� Setting: menu code
3�25�2
Congured in Setup under: Signum
®
2
Appl: tare. f: 3.25.
Restore factory default settings�
Configured in Setup under:
Appl: DEF.App: 9.1.
A number of subweighing operations
are required to form the basis for
calculation of an average weight�
You can enter the desired number
of subweighing operations using the
keypad�
The number you enter is active until
it is overwritten by another number
It also remains in memory when you
switch to a dierent application
program, or turn o the scale�
There are three ways to start the
averaging routine:
Manual start with preset number of
sub-weighing operations:
Place the sample on the platform and
press the O key
Manual start with user-dened number
of subweighing operations:
Place the sample on the platform
and enter the number of weighing
operations using the keypad� Press the
r key to save the number entered
and begin weighing
Automatic start with preset number of
subweighing operations:
Measurement begins when you place
the rst sample on the platform,
provided the start conditions are met�