64
Signum 3 Ex
Classification Application W
Classication
Application W
With the Classification application, you
can determine whether the weight of a
given sample lies within the limits of a
defined weight class�
Characteristics
Classification with 3 or 5 weight classes�
Configured in Setup under:
Appl 2:
class: 4.8.
Enter the upper limits of weight classes
using the keypad or by saving weight
values from a load on the platform
Enter the upper limits of weight classes
as absolute values or as a percentage
of deviation from the upper limit of
Class1 Congured in the menu under:
Appl 2:
class: 4.9.
Activate Info mode with the I key
Toggle the main display between
classication display and weight display
by pressing the w key
Automatic printout congured in Setup
under:
Appl 2:
class: 4.10.
Automatic taring of container weight�
Congured in Setup under:
appl: a.tare: 3.7.
Automatic initialization when you
switch on the scale with most recently
saved application data� Congured in
Setup under:
appl: a.start: 3.8.
You can assign dierent functions to
the c key for deleting applications�
When you clear applications, you can
delete either the data stored for all
applications or just selected data stored
for the active application� Congured in
Setup under:
Appl: sel.cf: 3.24.
(Sel� CF function key c in
applications)
Tare function:
1) If you store a tare (weight value)
by pressing the ) key, you can later
enter a tare value manuallyThe tare
value you enter is added to the stored
tare value� Setting: menu code 3�25�1
(factorysetting)
2) A tare value entered manually
overwrites a stored tare value
(weightvalue)�
If you enter a tare value manually, a
tare value (weight value) stored later
overwrites the manually entered value
Setting: menu code 3�25�2
Congured in Setup under:
Appl:
W:: 3.25.
Restore factory default settings�
Configured in Setup under:
appl: def.app: 9.1.
To use the Classication application, you
need to enter the delimiters that separate
one class from another
Limits between the individual weigh
classes are required for the classication�
The lower limit of Class 1 is dened by
the preset minimum load� The other
classes are congured by dening their
upper limits� There are two ways to enter
the delimiters for classes 1 through 3
(or5):
By saving the weight value indicated:
Each upper limit value, with the
exception of the highest class, is entered
using the keypad or by saving the weight
value of a load on the weighing platform�
By entering a percentage:
The upper value of Class 1 is entered
using the keypad or by saving the
value indicated� Upper limits for the
other classes are dened by entering a
percentage of deviation from the upper
limit of Class 1, using the keypad�
Example: Enter 100 g as the upper limit
of Class 1� Then enter 15%�
When working with 3 classes, this yields
the following weight classes:
Class 0: up to the minimum load
Class 1: >minimum load - 100 g
Class 2: >100 g – 115 g
Class 3: >115 g - maximum load
When working with 5 classes, this yields
the following weight classes:
Class 0: up to the minimum load
Class 1: >minimum load - 100 g
Class 2: >100 g – 115 g
Class 3: >115 g – 130 g
Class 4: >130 g – 145 g
Class 5: >145 g - maximum load
The initialization data remains valid
until deleted by pressing the c key or
until overwritten by a new value� They
remain saved after the scale is switched
o
Preparation
§ Turn scale on: Press the e key
§ While all segments are lit, press the
)key
§ Select Setup: Press the k key
repeatedly until appl is displayed
§ Conrm appl: Press the ) key
§ Select the Classication application:
Press the k key repeatedly and
conrm with the ) key
Classification Application Parameters
3�6� Minimum Load for Initialization
and Defining the Class 1 Lower
Limit
3�6�1* 1 digit
3�6�2 2 digits
3�6�3 5 digits
3�6�4 10 digits
3�6�5 20 digits
3�6�6 50 digits
3�6�7 100 digits
3�6�8 200 digits
3�6�9 500 digits
3�6�10 1000 digits
4�3� Control output for SET as:
4�3�1* Output “SET
4�3�2 Process control systems
ready to operate
4�7� Activation of outputs
4�7�1 Off
4�7�2 Always on
4�7�3* On at stability
4�8� Number of classes
4�8�1* 3 classes
4�8�2 5 classes
4�9� Parameter input
4�9�1* Weight values
4�9�2 Percentage values
4�10� Automatic printing
4�10�1* Off
4�10�2 On
* = Factory setting
§ Save the settings with the ) key and
exit Setup: Press the ( key several