Error Messages
Error Messages
Error messages are shown in the main display� Error messages are shown permanently� Information messages are shown for 2 seconds�
After this the program automatically returns to the weighing mode�
Display Solution
Err 101 Release key or
contact your local Minebea Intec Service
Center
Err 320 Contact your local Service Center
Err 340 Turn the scale o and then on again
If the error code is still displayed:
Contact your local Service Center
Err 341 Leave the device connected to power for
at least 10 hrs�
Err 343 Contact your local Service Center
Inf 01 Set output format correctly
inf 02 Unload the scale rst then zero
then tare using the ) key
inf 03 Allow to warm up again and repeat the
adjustment process
Inf 06 Contact your local Minebea Intec Service
Center
inf 07 Contact your local Service Center for
information on changing settings
inf 08 Check your conguration to ensure that
“Zero at power on” (1�12) is set�
Inf 09 Zero the scale�
inf 10 The data stored for the application
program must be deleted before
taring�
inf 22 Place a heavier weight on the platform
Inf 23 Contact your local Service Center
inf 29 Reduce min� load
(under Application, menu item 3�6)
inf 71 None
inf 72 None
Inf 73 Contact your local Service Center
inf 74 None
Inf 98 Contact your local Service Center
Inf 99 Contact your local Service Center
NO WP Contact your local Service Center
Flashing ! Setting the time
Inf 54/L
The load plate is not on the weighing
platform
Place the load plate on the platform
Inf 55/H
The load exceeds the capacity of the
weighing
platform
Unload the weighing platform
Care and Maintenance / Recycling
Cause
Key is stuck
Key pressed whilst switching on
Operating program memory faulty
Incorrect operating parameter (EEPROM)
RAM has lost data
Battery is empty
Loss of data in the memory area for trans-
action numbers in external alibi memory
Data output not compatible with output
format
Adjustment condition was not met
e�g� not tared or weighing pan loaded
Adjustment could not be completed within
a certain time
Built-in adjustment weight defective
Function not allowed in scales veried for
use in legal metrology
The load on the scale is too heavy to zero
the readout
Taring is not possible when the scale gross
weight is zero
Tare key is blocked when there is data in
the tare memory
Error in storing reference value,
weight is too low
Error in initializing an application
Minimum load not reached
Cannot store the current weight value
(e�g�, control limits too low or too high)
Cannot store the current weight value
(e�g� transaction counter maximum
reached)
Data not found or unreadable
Function is blocked
(e�g�, menu is locked)
No weighing platform connected
No weighing platform connected
No weighing platform connected
Battery defective or time changed