39
Signum 3 Ex
Operation
Operation
SIWX Models:
Internal Adjustment
Veried versions or with option E7
In the Setup menu (setup: :
wp-1:
1�9�) “internal adjustment” (setup:
wp-1: 1�9�4) must be congured�
setup
wp-1
1� 9� Calibration and
Adjustment
1�9�1 * Ext� calibration/
adjustment;
default weight
1�9�3 External calibra-
tion/adjustment;
weight can be
selected under
menu item 1�18�1
1�9�4 Internal calibra-
tion/adjustment
(models with
built-in motorized
calibration weight
only)
1�9�8 Set preload
1�9�9 Clear preload
1�9�10 J key locked
1�10� Calibration/
Adjustment
sequence
1�10�1 Calibration with
automatic adjust-
ment�
1�10�2 * Calibration with
manual adjust-
ment
* = Factory setting
The scale housing has a built-in motor-
ized calibration weight�
The calibration/adjustment procedure is
performed as follows:
§ Select calibration/adjustment: Press the
J key
> The internal calibration weight is loaded
automatically
> The scale is adjusted/calibrated
> In the Setup menu (
Setup: wp-1:
1�10�) If parameter 1�10�1 is selected in
Setup, the scale is adjusted automati-
cally
> In the Setup menu (
Setup: wp-1:
1�10�) If parameter 1�10�2 is selected in
Setup, the internal adjustment can be
stopped without adjusting the scale
> The internal adjustment weight is
unloaded from the scale
> ISO/GMP-compliant record: see page 91
Setting the Preload
Setup Information
! It is only possible to set a preload when
the menu access switch is open�
– The J key must be allocated to the
Set preload function (menu item 1�9�8)�
! After setting a preload, close the menu
access switch and reallocate the origi-
nal function back to the J key (e�g�
external calibration/adjustment with
user-dened weights) under menu item
1�9�
Clearing the Preload
Setup Information
! It is only possible to clear a preload
when the menu access switch is open�
The “Clear Preload" function must be
allocated to the J key
! After clearing a preload, close the
menu access switch and reallocate the
original function back to the J key
(e�g�external calibration/adjustment
with user-dened weights) under menu
item 1�9�