48
Signum 3 Ex
Counting Application Z
Preparation
§ Turn scale on: Press the e key
§ While all segments are lit, press the
)key
§ Select Setup: Press the k key
repeatedly until appl is displayed
§ Conrm appl: Press the ) key
§ Select the Counting application:
Press the k key repeatedly and
conrm with the ) key
Counting application parameters
3�6� Minimum load for initialization
3�6�1* 1 digit
3�6�2 2 digits
3�6�3 5 digits
3�6�4 10 digits
3�6�5 20 digits
3�6�6 50 digits
3�6�7 100 digits
3�6�8 200 digits
3�6�9 500 digits
3�6�10 1000 digits
3�9� Resolution for
calculation of reference value
3�9�1* Display accuracy
3�9�2 Display accuracy + 1
decimal place
3�9�3 Display accuracy + 2
decimal places
3�9�4 Internal resolution
3�11 Parameter for saving weight values
3�11� 1* With stability
3�11� 2 With increased
stability
3�12� Average piece weight updating
3�12�1 O
3�12�3* Automatic
* = Factory setting
§ Save the settings with the ) key and
exit Setup: Press the ( key several
times�
Parameter for Saving Weight Values
The weight on the platform is saved as
a reference value when the platform
has stabilized� “Stability” is defined as
the point at which the fluctuation of
a measured value lies within a defined
tolerance range� The narrower the
tolerance range, the more stable the
platform is at “stability”�
In Setup menu:
appl1: count: 3.11.
You can dene whether the value is
saved when “standard stability” is
reached, or only at “increased stability”
(narrower tolerance range)� If you
select “increased stability,” the value
saved for average piece weight will be
more accurate and the results more
reproducible, but the response time of
the weighing platform might be longer�
Accuracy of Average Piece Weight
Calculation
The resolution applied for calculating
the reference weight is dened in Setup
under:
Appl1
: count: 3.9.
The resolution for calculating the
reference weight is increased if “+1
decimal place,” “+2 decimal places” or
“Internal resolution” is selected� With
the “+1 decimal place” setting, the net
value is determined to one additional
decimal place (i�e�, display accuracy
x 10); “+2 decimal places” increases
display accuracy + 100, and so on up
tothe maximum resolution available�
Minimum Load
The minimum load required for
initialization of the weighing platform
is configured in Setup under:
appl1:
count: 3.6.
Once the limit is exceeded by the load,
initialization can begin�
If the load on platform is too light,
thefollowing will occur when you try
tosave a value:
– Error code Inf 29 appears,
– The weighing platform is not initialized
– The preset reference sample quantity is
saved
The minimum load required for
automatic taring of the container
weight on the platform (rst weight)
iscongured in Setup under:
appl
: m.weigh: 3.5.
You can choose from the following
10levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform� If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you
must place at least 1000 g (= 1000
intervals) on the weighing platform for
initialization�
Counting Application Z