86
8 Error Displays and Countermeasures
No. 99MAH056A
No. Error name LCD display Cause Countermeasures
51
Calibration
Schedule
Warning Forced
Error
Err 51
A serial
communication
command [89]
(calibration
schedule warning
forced display) is
received.
Send a serial
communication
command [89]
(calibration schedule
warning forced display).
61
Set Value
Rewrite Error
Err 61
For some reason,
the set value has
been rewritten
since the previous
use.
Press the [F1] key to
return to measurement
mode, check various
settings, and reset where
needed.
62
Set Value
Storage Error
Err 62
Settings cannot be
saved.
The set values
cannot be read.
y After turning OFF the
power, turn ON the
power again, check
various set values,
and reset where
needed.
y If the same error
occurs even after the
power is turned on
again, the product
may be faulty. Please
contact the agent
where you purchased
the product or a
Mitutoyo sales/service
representative.
(
"12O󰀨-Site
Repairs (Subject
to Charge)" on
page 105)
y If the error occurs
frequently, the supply
voltage may be
unstable. Check the
supply voltage.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mitutoyo ID-C Series and is the answer not in the manual?

Mitutoyo ID-C Series Specifications

General IconGeneral
BrandMitutoyo
ModelID-C Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals