Page #99 background image
87
No. 99MAH056A
8 Error Displays and Countermeasures
No. Error name LCD display Cause Countermeasures
63
Internal Program
Error
Err 63
Measurement
cannot be done
because of an
internal program
error.
The product may be
faulty. Please contact
the agent where you
purchased the product or
a Mitutoyo sales/service
representative.
(
"12O󰀨-SiteRepairs
(Subject to Charge)" on
page 105)
90
Allowable Value
Setting Error
Err 90
The upper limit
value is set to a
value smaller than
the lower limit
value.
Set the upper limit value
to a value greater than
the lower limit value
(upper limit value > lower
limit value).
(
"5.1 Selecting
Parameter Items" on
page 35,
"5.6.2
Setting Allowable Values
(Upper Limit Value and
Lower Limit Value)" on
page 48)
91
Calculation
Coe󰀩cient
Setting Error
Err 91
The calculation
coe󰀩cientisset
to 0.
Reset the calculation
coe󰀩cienttoavalue
other than zero.
(
"5.1 Selecting
Parameter Items" on
page 35,
"5.7
Setting Calculation
Function Selection and
CalculationCoe󰀩cient"
on page 50)
92
Calibration Date
Setting Error
Err 92
The calibration
date and advance
warning date are
set before the
current date.
Set it so that current date
< advance warning date
< calibration date.
(
"5.1 Selecting
Parameter Items" on
page 35,
"5.11. 2
Calibration Schedule
Warning Selection/
Setting" on page 66)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mitutoyo ID-C Series and is the answer not in the manual?

Mitutoyo ID-C Series Specifications

General IconGeneral
BrandMitutoyo
ModelID-C Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals