82BAttribute Summary 11BAppendix C: Command Summary
100
Table 51: Attribute Summary (continued)
Attribute
ID #
Class
Code (hex)
Object
Model (dec) (hex)
Description
DeviceNet Data
Type
Access
Rule
Non-
Volatile
Memor
y
Factory
Default
Setting
03 0x03 Data Type USINT U-Set/Get Y 0xC3
(INT)
04 0x04 Data Units ENGUNITS U-Set/Get Y 0x1001
(Counts)
05 0x05 Reading Valid BOOL U-Get N none
06 0x06 Value INT or REAL U-Get N none
07 0x07 Status Byte U-Get N none
08 0x08 Alarm Enable BOOL U-Set Y 0 (disable)
09 0x09 Warning Enable BOOL U-Set Y 0 (disable)
10 0x0A Full Scale INT or REAL U-Get Y 23405
14 0x0E Gain REAL U-Set/Get Y 1.000
16 0x10 OffsetB Customer Offset INT or REAL U-Set/Get Y 0
17 0x11 Alarm Trip Point High INT or REAL U-Set Y 50% FS
18 0x12 Alarm Trip Point Low INT or REAL U-Set Y 50% FS
19 0x13 Alarm Hysteresis UINT or REAL U-Set Y 0
20 0x14 Alarm Settling Time UINT U-Set Y 0
21 0x15 Warning Trip Point
High
INT or REAL U-Set Y 50% FS
22 0x16 Warning Trip Point Low INT or REAL U-Set Y 50% FS
23 0x17 Warning Hysteresis UINT or REAL U-Set Y 0
24 0x18 Warning Settling Time UINT U-Set Y 0
31 0x1F Averaging Time UINT U-Set/Get Y 30
32 0x20 Overrange INT or REAL U-Get Y 110% FS
33 0x21 Underrange INT or REAL U-Get Y -5% FS
94 0x5E Sensor Warning STRUCT of
BYTE BYTE
U-Get N none
95 0x5F Sensor Alarm STRUCT of
BYTE BYTE
U-Get N none
96 0x60 Status Extension BYTE U-Get N none
99 0x63 Sub Class UINT U-Get Y 3
0x31
S-Analog
Sensor
119 0x77 Pressure Fraction in
Percentage Full Scale
REAL U-Get Y none
(Continued on next page)