62 | P a g e
Green Button is bad
Replace green button
Alarm Screen Say “Formalin
Dispense Low” at bottom of
screen
Internal Formalin storage
Tank is low
Add formalin to internal tank using transfer pump
Formalin Dispensing Leak alarm is
present (sensor #1, inside station,
beneath formalin storage tank)
There is a formalin leak or
sensor is faulty
Repair leak & clean sensor of fluid or replace faulty
sensor
FORMALIN COLLECTION SYSTEM
Formalin Collection Full alarm
Formalin collection
Jug/Carboy is full
Empty formalin Jug/Carboy
Float switch disconnected
Plug float switch into station.
Check wires in connector are not pulled out
Float switch “bulb” is stuck in
the “full” (up) position
Unscrew cap on Jug/Carboy and make sure metal
“bulb” on float switch can move up and down the
rod without getting stuck. Or replace float switch.
Keeping the system full reduces the
Keeping the system full helps eliminate buildup on
the float sensors.
Formalin Collection Leak alarm is
present (sensor #2 (sensor #2,
located in formalin collection
cart)
Sensor disconnected
Plug sensor into station.
Check wires in connector are not pulled out
There is an active leak/spill in
the bottom of the cart
Repair leak & clean sensor of fluid or replace faulty
sensor
Alarm Buzzer Notification
Alarm buzzer is ON
Check alarm screen at the
bottom of the screen for
alarm identification or Press
the Alarms button on the
Action page.
Identify alarm cause and correct the alarm
condition.
Press Alarm Setting button
on the Actions page
Identify alarm cause and correct the alarm
condition. Verify which alarm condition has
audible notification and set them according to
Laboratory preferences.