SUPREMATouch
68
Operation of the System
MSA
US
Fig. 39 “Measure Points”
The functions of the individual fields are described below:
Measure Point No.:
Field type: Selection
After selection of a measure point number, the current signals of the selected point are displayed.
Signal UA:
Field type: Display
The amplified sensor signal is displayed in these fields separated by bus.
Signal UQ:
Field type: Display
When passive sensors and MPI modules are used, the bridge current is displayed as a voltage
value in these field separated by bus. When active sensors and MCI modules are used, these
fields are empty.
Signal UY:
Field type: Display
When passive sensors and MPI modules are used the amplified sensor signal UY is displayed in
these fields separated by bus. The signal consists of a fixed gain that depends on the MPI module
type used and an offset voltage.
When active sensors and MCI modules are used, these fields are empty.
Calc. UX + offset:
Field type: Display
These fields show the calculated UX value including an offset. These values are calculated based
on the measured UY values.