EasyManua.ls Logo

MTE HYDROCAL1005 - 5.10.2 External sensors (see also chapter 4.7.3)

MTE HYDROCAL1005
113 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HYDROCAL 1005 Manual for Installation and Operation Page 84/111
Selection is made by the menu and with right mouse click
or by „View->View As->...“
In the views „List“ and „Symbol“, the following dialogue can be opened by double clicking on one of
the alarms (in view „Details“ it is not possible to open this dialogue):
First the mode of alarms must be selected under „Alert Mode“. Following modes are available:
High: An alarm is triggered when Level 1 is exceeded.
High-High: An alarm in two steps is triggered. The 1
st
alarm is triggered when Level 1 is ex-
ceeded; the 2nd alarm is triggered after exceeding Level 2.
Low: An alarm is triggered when a value falls below the limit defined in Level 1.
Low-Low: An alarm in two steps is triggered. Step 1 is alarm is triggered when the measured
value falls below the limit defined in Level 1; the second alarm is triggered when the measured
value falls below the limit defined in level 2.

Table of Contents

Related product manuals