Calibrating an LVDT
Model 493.02 Controller Service
Calibration
101
5. Set conditioner polarity to Normal (default) so that a positive
command extends the actuator.
Stand-Alone — Path: Setup > Disp > Calibration > Polarity >
Normal
Automated — On the Calibration tab, select Normal for
conditioner Polarity.
6. Check to ensure that the polarity of the servovalve signal is set
correctly. See “Setting the Servovalve Polarity” on page 73.
Stand-Alone — Path: Setup > Output > Polarity
Automated — Click the Channel Drive icon to open the
Channel Drive panel, then click the Valve tab.
Check the valve polarity setting. If you do not know what polarity
to use for your servovalve, select Normal (default).
7. Set P Gain to 0.5
Stand-Alone — Path: Tuning > P Gain
Automated — From Station Setup, click the Channel Tuning
icon, click the Adjustments tab, and then set P Gain.
8. Ensure that displacement control mode is selected.
Stand-Alone — Path: Status > Control Mode > Disp.
Automated — Check Control Mode on the Station Manager
Function Generator panel.
9. Ensure that the Span is set for 100%.
Stand-Alone — Path: Status > Span
Automated — Check the Master Span setting on the Station
Controls panel

Table of Contents

Other manuals for MTS Systems FlexTest SE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest SE and is the answer not in the manual?

MTS Systems FlexTest SE Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest SE
CategoryController
LanguageEnglish

Related product manuals