EasyManua.ls Logo

MTS Systems FlexTest SE - Task 6 Gain;Delta K Calibration

Default Icon
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 493.02 Controller Service
Calibrating an LVDT
Calibration
114
Task 6 Gain/Delta K Calibration
If you using Gain/Delta-K for your calibration type, complete the
following procedure. If not, complete Task 7 Gain/Linearization
Calibration on page 119
Calibrate actuator
retraction
LVDTs can be
calibrated so that a
positive output
represents actuator
extension and a
negative output
represents actuator
retraction, or vice
versa.
You calibrate the
negative side of the
output with gain and
the positive side of
the output with Delta
K.
Delta K compensates
for differences in symmetry between the positive and negative outputs.
You should calibrate actuator extension at 80% of the full scale range.
Gain controls Pre-Amp gain is a selectable gain amplifier with predefined values.
Since changes in Pre-Amp gain can cause spikes in the feedback
signal, Pre-Amp gain can only be adjusted when hydraulics are off.
Post-Amp gain is a finer, operator-defined gain control that can be
adjusted when hydraulics are on.
The Total Gain value is calculated by multiplying the Pre-Amp and
Post-Amp gain values. If the total desired gain amount is known (from
a calibration sheet), you can enter the amount in the Total Gain box
and the software will calculate the Pre-Amp and Post-Amp gain
values automatically.
When you are calibrating an AC conditioner, use the Post-Amp gain
control to increase gain. If more gain is needed, you must disable
mechanical input
Indicated output
K > 1
K = 1
K <
1
Gain

Table of Contents

Other manuals for MTS Systems FlexTest SE

Related product manuals