Calibrating an LVDT
Model 493.02 Controller Service
Calibration
119
Task 7 Gain/Linearization Calibration
If you are using Gain/Linearization for your calibration type,
complete the following procedure. If not, complete Task 6 Gain/Delta
K Calibration on page 114.
Important Using linearization data requires specific conditioner zeroing
practices. Ensure that Electrical Zero Lock on the Offset/
Zero menu is set to Locked. Adjusting electrical zero after
calibration may invalidate linearization data.
Initial LVDT
calibration
For initial calibration of an LVDT complete the following procedure:
1. Select Gain/Linearization for Cal Type.
Stand-Alone— Path: Setup > Disp. > Calibration > Cal Type >
Gain/Linearization
Automated— Select Gain/Linearization for Cal Type on the
Calibration tab of the Inputs panel.
2. Apply a retraction command that is 80% of the full scale range.
Stand-Alone:
A. Enable the front panel Dial.
B. Use the Dial to adjust the Setpoint for 80% of the full scale
range.
C. Verify that your LVDT displacement signal equals 80% of the
full scale range.
Automated:
A. Open the Manual Command window, and then click on the
Enable Manual Command.
B. Adjust the Manual Cmd slider on the Manual Command
window for 80% of the full scale range.
C. Use the Station Signals panel to verify that your LVDT
displacement signal equals 80% of the full scale range.
During the initial calibration and tuning of your system, it may
take repeated adjustment for the two values to match.

Table of Contents

Other manuals for MTS Systems FlexTest SE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest SE and is the answer not in the manual?

MTS Systems FlexTest SE Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest SE
CategoryController
LanguageEnglish

Related product manuals