Model 493.02 Controller Service
Calibrating an LVDT
Calibration
126
E. Record the dial indicator and conditioner feedback values at
the -2% row of your record sheet.
F. Repeat steps D and E for other retraction commands
(typically at -4, -6, -8, -10, -20, -40, -70, and -100 percent of
full scale).
11. Record dial indicator and conditioner feedback readings at
predetermined extension command points.
Stand-Alone:
A. Use the front panel Dial to adjust the Setpoint for a +2%
extension command.
B. Record the dial indicator value and conditioner feedback at
the +2% row of your record sheet.
C. Repeat steps A and B for other extension commands
(typically at +4, +6, +8, +10, +20, +40, +70, and +100 percent
of full scale).
Automated:
A. Adjust the Manual Cmd slider for a +2% extension
command.
B. Record the dial indicator value and conditioner feedback at
the -2% row of your record sheet
C. Repeat steps A and B for other extension commands
(typically at +4, +6, +8, +10, +20, +40, +70, and +100 percent
of full scale).
12. Turn off system hydraulics.

Table of Contents

Other manuals for MTS Systems FlexTest SE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest SE and is the answer not in the manual?

MTS Systems FlexTest SE Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest SE
CategoryController
LanguageEnglish

Related product manuals