Model 493.02 Controller Service
Calibrating a Force Sensor
Calibration
148
7. Reset Linearization Data window to default values.
Stand-Alone— Path: Setup > Force > Calibration > Cal Type >
Gain/Linearization > <<Reset>>
Automated— Click Reset on the Linearization Data window to
return to default values.
8. Exercise the force standard.
Stand-Alone— Enable the front panel Dial, then adjust the
Setpoint to cycle the load standard readout between zero and full
tension three times. This removes sensor hysteresis. Path: Status >
Setpoint
Automated— Use the Manual Cmd slider on the Manual
Command window to cycle the load standard readout between
zero and full tension three times. This removes sensor hysteresis.
Example: When calibrating a ±10 kN range, exercise the actuator
between 0 and -10 volts. To calibrate the same force sensor for a
different range such as ±5 kN, exercise the load standard between
0 and -5 volts.
9. Apply a tensile force command that is 80% of the range’s full
scale.
Stand-Alone:
A. Using the Dial, adjust the Setpoint for a tensile force
command that is 80% of the full scale range.
B. Verify that your force transducer feedback signal is 80% of the
full scale range.
Automated:
A. Adjust the Manual Cmd slider for a tensile force command
that is 80% of the full scale range.
B. Verify that your force transducer feedback signal is 80% of the
full scale range.
During the initial calibration and tuning of your system, it may
require repeated adjustment for the tensile force command and
feedback values to match.