Model 493.02 Controller Service
Calibrating a Force Sensor
Calibration
154
Where:
Conditioner Output Voltage is typically 10 Vdc.
Excitation Voltage is specified using the Excitation entry box on
the Calibration submenu or tab.
Compression is specified using the Negative Compression or
Positive Compressive entry box on the Calibration submenu or
tab.
mv/v Positive Tension
Calibration
Use the following procedure if your force transducer is set up so that a
positive output represents actuator retraction (tension).
1. Select mv/v Pos Tension for your calibration type.
Stand-Alone— Path: Setup > Force > Calibration > Cal Type >
mv/v Pos Tension
AutomatedSelect mv/v Pos Tension for Cal Type on the
Calibration tab of the Inputs panel.
2. From the Calibration Data sheet for your force transducer enter
the following values on the Calibration submenu or tab:
Stand-Alone:
A. Enter the full scale force values in the Minimum and
Maximum entry boxes.
B. Enter the tension sensitivity value (+mV/v) in the Pos
Tension entry box.
C. Enter the compression sensitivity value (-mv/V) in the Neg
Compression entry box.
D. Enter the calibration excitation value (Vdc) in the Excitation
entry box.
Gain
Conditioner Output Voltage
Excitation Voltage Compression mV/V×
-----------------------------------------------------------------------------------------------------------
=

Table of Contents

Other manuals for MTS Systems FlexTest SE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest SE and is the answer not in the manual?

MTS Systems FlexTest SE Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest SE
CategoryController
LanguageEnglish

Related product manuals