Model 493.02 Controller Service
Calibrating a Force Sensor
Calibration
156
Automated:
A. Enter the full scale force values in the Fullscale Min/Max
entry boxes.
B. Adjust Neg Tension for the required tension sensitivity value
(+mv/v).
C. Adjust Pos Compression for the required compression
sensitivity value (-mv/V).
D. Adjust Excitation for the required calibration excitation value
(Vdc).