Model 493.02 Controller Service
About Tuning
Tuning
198
CLC control mode Channel limited channel (CLC) control modes are used for specimen
installation and removal. Channel limited channels require two
feedback signals. See “Tuning a CLC Control Mode” on page 219
Refer to “Tuning” in Chapter 3, Menus Descriptions of the FlexTest SE
Users Manual for more information about CLC tuning controls and
guidelines.
Dual Compensation
control mode
Select a dual compensation control mode for a channel when the
feedback for the desired control mode is unsuitable for maintaining
closed-loop control.
A dual compensation mode requires a primary and a secondary feedback
signal:
• The more stable primary feedback signal is used by the PIDF
controller to maintain closed-loop control.
• The less stable secondary feedback signal is used for command
compensation in command programs provided by the Function
Generator and other applications.
Refer to “Tuning” in Chapter 3, Menus Descriptions of the FlexTest SE
Users Manual for more information about dual compensation tuning
controls and guidelines.