Power-Up and Testing 4-31
! The Dispenser I/O Test dialog box (Figure 4-20) opens.
Figure 4-20 Dispenser I/O (Inputs 0-31)
4. Locate the bit labeled TACTILE_SENSOR.
5. Lightly press on the Tactile Needle Sensor.
! The
TACTILE_SENSOR bit should toggle to ON (1) when the Tactile Needle Sensor is
pressed and return to
OFF (0) when pressure is released.