EasyManuals Logo

Nordson Asymtek Spectrum S-920 User Manual

Nordson Asymtek Spectrum S-920
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #97 background imageLoading...
Page #97 background image
Power-Up and Testing 4-31
! The Dispenser I/O Test dialog box (Figure 4-20) opens.
Figure 4-20 Dispenser I/O (Inputs 0-31)
4. Locate the bit labeled TACTILE_SENSOR.
5. Lightly press on the Tactile Needle Sensor.
! The
TACTILE_SENSOR bit should toggle to ON (1) when the Tactile Needle Sensor is
pressed and return to
OFF (0) when pressure is released.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nordson Asymtek Spectrum S-920 and is the answer not in the manual?

Nordson Asymtek Spectrum S-920 Specifications

General IconGeneral
BrandNordson
ModelAsymtek Spectrum S-920
CategoryDispenser
LanguageEnglish

Related product manuals