NORMALAB NDI 450 - Page 112

NORMALAB NDI 450
118 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NORMALAB France
2012
N O R M A L A B F r a n c e
112/118
Alarm
Action level 1
Action level 2
Action level 3
Drip-deflector
missing :
check presence of drip deflector
drip deflector must be correctly
adjusted
contact technical services
Receiver
chamber door
open:
check if door is correctly closed
check
contact technical services
Condenser bath
liquid missing:
check the level in the tank
add coolant if the level is too low
contact technical services for
leakage control
Receiver volume
incorrect :
check receiver position, cleanliness and volume
check or review volume calibration
contact technical services
Receiver missing :
check presence of the receiver in the compartment
check receiver positioning
contact technical services
Sample probe
problem :
check probe connection
check sample probe positioning
contact technical services
Receiver
chamber probe
problem :
check that compartment temperature is between 0 &
100°C
check probe positioning
contact technical services
Condenser bath
probe problem :
check that condenser temperature is between 0 &
100°C
contact technical services
Ambient probe
problem :
Check Probe position
contact technical services
Heating probe
contact technical services