Solution or check to perform
The circuit breaker does
not remain engaged.
Verify that the mains voltage is in
concordance with that displayed on the
instrument.
The circuit breaker is
tripped when a test is
started.
Check that the transformer of the flask
heating circuit.
Switchover the switch on the two internal
power supplies and verify whether the
problem persists.
The fuse of the heating
block heating element
has blown.
Verify that the heating element circuit is not
damaged.
The results are
divergent in
temperature
Verify that the sample temperature probe is
correctly displayed and calibrated.
Verify that calibration of the test tube and its
declaration.
Verify that calibration of the barometric
sensor.
Verify that the final volume.
Verify that leak tightness of the stopper
connection between the flask and condenser
tube.
Initial boiling point is
divergent
Verify that temperature conditioning of the
sample on starting the test.
Verify that the position of the probe in the
flask neck (see standard).
Verify that the power leads for the internal