4.2 Options and equipment required
Sensor to be
calibrated
By
simulation
By
bath
By
point
Equipment necessary
Simulator + cable
reference XXX
Bath + extension lead
reference 60525 +
and the sensor to be
calibrated
Life book with
previous values
Sample probe
Possible
Cable ref 60522
Recommended
Possible
Condenser probe
Recommended
Cable ref 60524
Difficult to perform
Possible
Compartment probe
Recommended
Cable ref 60524
Difficult to perform
Possible
Ambient probe
Recommended
Cable ref 60524
Difficult to perform
Possible
TC dry point
Recommended
Cable ref 60523
Difficult to perform
Possible
TC Plate
Recommended
Cable ref 60523
Difficult to perform
Possible
The instrument does not allow storing in memory more than one calibration for fixed probes
(condenser, compartment and ambient probes).
For the sample probe, subject to much handling, it is possible to store as many calibrations as probes
are required. As a result with these previous calibrations, if the sample probe is defective or broken,
it is possible to connect another probe which has been previously calibrated and to select this
calibration to activate it.
4.3 Calibration of sample temperatures