7 Initial Operation
16 / 86 Instruction Manual Mesotest II (G/62800/0715/en)
7 Initial Operation
7.1 Before Initial Operation
Before the Mesotest II is operated for the first time, you must complete
the following actions:
Set up
Connection
If necessary, install the software
Attention
Risk of incorrect measurements/equipment damage due to improper set-
up
Before first use, make sure the installation and connection of the
Mesotest II are completed by our service or by a professional
authorized by OCULUS.
7.2 Information About the Room in which the Device is Set Up
The Mesotest II should be set up in a moderately bright or darkened room
(room brightness between 2 and 5 lux).
Attention
Interruption of the examination due to overheating
If the unit overheats, it shuts itself off and the test results are lost.
Do not cover the ventilation openings.
Note:
Falsified examination results if the room is too bright
If you set up the Mesotest II in a bright room, you must occlude the back
pane of the viewer with the supplied cover.
Insert the cover, see sec. 7.2.1, page 17