20 Appendix
Instruction Manual Mesotest II (G/62800/0715/en) 83 / 86
To be in compliance with the requirements of the IEC 60601-1-2. 6.1 and
6.2 the following types of equipment, accessories, power adapters and
cables must be used.
20.2 Guidance and Manufacturer’s Declaration -
Electromagnetic Emmisssions for the Mesotest II
Order number Description
62800
62801
Mesotest II
Mesotest II konkav
05200320 Kabel with plug, EU standard 2.5 m
(8.2021 feet)
05200210 (110 Volt) Kabel with plug, US standard 2.5 m
(8.2021 feet)
Guidance and manufacturer´s declaration electromagentic emmissions
IEC 60601-1-2, 5.2.2.1, table 1
The OCULUS Mesotest II is intended for operation in the electromagnetic environment specified below. The user
of the Mesotest II should ensure that it is being used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The Mesotest II uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
HF-emissions
CISPR 11
Class B
Harmonics emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
complies