13 Cleaning, Disinfection and Maintenance
Instruction Manual Mesotest II (G/62800/0715/en) 69 / 86
Unplug the power cable from the socket (1).
Push the button at the control pad plug and pull it out of the jack
(2), or unplug the connecting cable to the computer/laptop.
13.2 Cleaning
Required Materials
Cleaner for plastic surfaces with anti-static effect
Cleaner for painted surfaces: Mixture of equal parts of alcohol and
distilled water, possibly with a few drops of commercial detergent
Soft cloth or lens brush
Alcohol or lens cleaner
Mild soap solution
Cleaning intervals: Wipe the viewer enclosure after each examination,
and the housing whenever necessary.
Turn the Mesotest II off, sec. 7.5, page 19.
Unplug the power cord.
When cleaning, use a damp cloth and make sure that no liquid enters
the Mesotest II.
Clean the plastic surfaces and painted surfaces with appropriate
cleaning agents.
Clean the lenses with a soft cloth or lens brush, and, if necessary,
with alcohol or a lens cleaner.
Clean the viewer enclosure with a mild soap solution.
13.3 Disinfection
Required Materials
To disinfect all surfaces (except the Plexiglas back pane of the viewer) we
recommend the use of:
Mikrozid sensitive wipes premium
Make: Schülke & Mayr GmbH; Softpack 48x, Art.No. 165711
Disinfect the viewer enclosure after every examination and the
housing whenever necessary.
If the viewer enclosure is very dirty, you can exchange it, sec. 13.4.1,
page 71.
Note:
Equipment damage caused by disinfectant solution
The disinfectant solution may damage the finish if it is sprayed directly
on it.
Spray the disinfectant solution onto a cleaning cloth, do not spray it
directly on the device.