EasyManua.ls Logo

OCULUS Mesotest II - Transport and Storage; Transport and Storage Information

OCULUS Mesotest II
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15 Transport and Storage
Instruction Manual Mesotest II (G/62800/0715/en) 75 / 86
15 Transport and Storage
The Mesotest II, must be properly dismantled and packed before being
transported or stored.
15.1 Transport and Storage Information
Storage
This device does not keep up with the temperature conditions for storage
according to ISO 15004-1.
Do not store the Mesotest II under conditions over 50 °C and under
-10 °C.
In accordance with DIN EN 60601-1, the storage conditions are:
Transport
In accordance with DIN EN 60601-1, the transport conditions are:
After storage and transport
Wait approx. 3-4 hours after transport before operating the
Mesotest II for the first time. In the event of extreme temperature
changes from cold areas to warm rooms, the optical components can
become fogged.
Ambient temperature range -10°C to +55°C
Relative humidity, including condensation 10% to 95%
Air pressure range 700 hPa to1060 hPa
Ambient temperature range -40°C to +70°C
Relative humidity, including condensation 10% to 95%
Air pressure range 500 hPa to 1060 hPa

Table of Contents

Related product manuals