14 Messages, warnings and error messages / fault correction
14.1 Information and warnings
Surgeon name must be unique
Choose a surgeon name with fewer letters.
Connect a pedal or switch pedal on.
Phaco handpiece defective
Plug in another phaco handpiece or a handpiece.
- Check whether tip is broken.
- Tighten tip correctly.
- Check whether air pressure is on.
- Plug in air hose.
Wait for a short time while machine can cool down.
Turn the key switch for the laser on.
Laser emergency switch off
Release emergency switch.
Laser: no fibre detection - fully connect fibre.
Insert fibre into laser port.
Laser door interlock not closed
Remind: Laser user protection
Check whether laser protection filter in microscope is in place and all
persons in the room wear their laser protection glasses.
No serial number on print
Empty cassette into a drainage bag or change cassette.
- Change HF tip or HF handpiece.
- Decrease laser power and/or pulse duration.
- Increase interval between two pulses.
Laser: output power too low
- Check fibre and recalibrate laser.
- Call for laser second level support.
Call for laser second level support.
Laser microphone malfunction
Call for laser second level support.
- Wait until system has cooled down.
- Reduce laser power or increase pulse interval.
Laser: no auto key detected
- Auto key at the laser module (connector at front of laser module) is
missing. Call service.
Laser: foot switch malfunction
Call for laser second level support.
Laser: no foot switch detected
- Connect OS 4 foot pedal and confirm.
- Switch pedal on and confirm.
Laser: Active user protection filter detected
although it is not configured. Active user
protection filter is used.
If active user protection filter is installed permanently, change the config-
uration for the user protection filter in the service mode to “active”.
Laser: active user protection filter is plugged
but not in place.
Install the active UPF in the microscope according to the manufacturer's
instructions.
Laser: no active user protection filter de-
tected – connect active user protection filter
Connect active user protection filter.
If you wish to permanently use a passive user protection filter, install a
passive user protection filter and change the configuration for the user
protection filter in the service mode to “passive”.