2021-03-23 / Rev. 10 VV016042 129
ERROR 268
Laser: user protection filter malfunction
- Press «OK, disable LASER» to disable the laser
function. All other functions of the OS 4 remain availa-
ble.
- Call for laser second level support
ERROR 269
Laser: beam switch malfunction
- Press «OK, disable LASER» to disable the laser
function. All other functions of the OS 4 remain availa-
ble.
- Call for laser second level support
ERROR 273
Laser: I2C IO expander malfunction
- Press «OK, disable LASER» to disable the laser
function. All other functions of the OS 4 remain availa-
ble.
- Call for laser second level support
ERROR 274
Laser: processor malfunction
- Press «OK, disable LASER» to disable the laser
function. All other functions of the OS 4 remain availa-
ble.
- Call for laser second level support
15 Authorized service points
Oertli Instrumente AG
Hafnerwisenstrasse 4
Customer service and support
CH 9442 Berneck, Switzerland
Tel.: +41 (0)71 747 42 00
E-mail: css@oertli-instruments.com
Website: www.oertli-instruments.com
Contact your local distribution partner or Oertli Instrumente AG for information on service points in your country authorized
by Oertli.
15.1 Sending defective device, instrument or part to Oertli
1. Clean and sterilise instruments before returning them to the service point.
2. Fill out the QMF 17.09E and QMF 17.15E
3. Send device, instrument or part in appropriate packaging to the service point.