Functional Test
OS 4
VV016042, 2021-03-23 / Rev. 09
Open I/A tubes and hold them into an empty vessel.
Notice: _____________________________________________________________________________
Activate reflux and generate occlusion.
Notice: _____________________________________________________________________________
Press „Continuous Irrigation“ key and “BSS extraction” key.
Irrigation is retained (liquid is
flowing from irrigation tube)
Notice: _____________________________________________________________________________
Activate „BSS extraction“ key.
Notice: _____________________________________________________________________________
Assemble phaco handpiece with phaco tip and test chamber and install tubing system!
Select phaco, press „TEST“ key.
Notice: _____________________________________________________________________________
After completion of the phaco test, the notice „Test OK“ appears.
Notice: _____________________________________________________________________________
Activate maximum phaco power output.
Notice: _____________________________________________________________________________
Activate pulse, burst and CMP.
Notice: _____________________________________________________________________________
Generate occlusion with surgeons‘ values and undo occlusion.
Test chamber remains stable
Notice: _____________________________________________________________________________
No HF instrument is connected!
Select HF / Dia and press „TEST“ key.
Notice: _____________________________________________________________________________
After completion of the HF test, the notice „TEST OK“ appears.
Notice: _____________________________________________________________________________
Connect HF instrument and immerse any HF tip in liquid!
Function OK, sound correct
Notice: _____________________________________________________________________________