130
Chapter 6 Reprocessing, Storage, and Disposal
HIGH FLOW INSUFFLATION UNIT UHI-4
Dry the Insufflation Tube (MAJ-590) and Suction Tube
(MAJ-591)
If disinfect the accessories soon after, this step may be skipped.
1. Wipe off any water droplets on the external surfaces of the tubes and
connectors with sterile lint-free cloths.
2. Using compressed air to remove water droplets inside the tube in advance
or using a drying cabinet for surgical instruments is useful to reduce the
leaving time required to dry.
• The leaving time required to dry completely depends on the
temperature and humidity of the ambient environment.
• Using compressed air to remove water droplets inside the
tube in advance or using a drying cabinet for surgical
instruments is useful to reduce the leaving time required to
dry.