130
Chapter 6 Reprocessing, Storage, and Disposal
HIGH FLOW INSUFFLATION UNIT UHI-4
Dry the Insufflation Tube (MAJ-590) and Suction Tube
(MAJ-591)
If disinfect the accessories soon after, this step may be skipped.
1. Wipe off any water droplets on the external surfaces of the tubes and
connectors with sterile lint-free cloths.
2. Using compressed air to remove water droplets inside the tube in advance
or using a drying cabinet for surgical instruments is useful to reduce the
leaving time required to dry.
The leaving time required to dry completely depends on the
temperature and humidity of the ambient environment.
Using compressed air to remove water droplets inside the
tube in advance or using a drying cabinet for surgical
instruments is useful to reduce the leaving time required to
dry.