40
Chapter 3 Installation and Connection
HIGH FLOW INSUFFLATION UNIT UHI-4
When attaching the cylinder hose to the CO
2
gas inlet on the
rear panel of the UHI-4, unbend the cylinder hose and tighten
the hose holding by hand (see Figure 3.2). Otherwise, there
is a fear of significant gas leakage.
Figure 3.2
Cylinder hose for UHI-3 is sold separately. Please purchase
a cylinder hose for UHI-3 compatible with your gas cylinder
type. For specific details, please contact Olympus.
Attach a cylinder holder on an Olympus trolley to secure the
CO
2
cylinder before use. For installation, also refer to the
Instructions manual for the cylinder holder used.
The dedicate trolley (WM-NP2/WM-DP2) is required to use
this device in combination with the cylinder hose with switch
over valve (MAJ-1985, MAJ-1986). When using the cylinder
hose with switch over valve, refer to the instruction manual.
For EU, Australian and New Zealand market
Cylinder hose must be attached to UHI-4 by Olympus personnel on installation
time. Enclosed nut cover will be fixed at the place where cylinder hose and
insufflator are connected by Olympus personnel.
Trolley Applicable cylinder holders
T1/TC series MH-509, MH-510, MH-511
WM-NP1, WM-WP1, WM-DP1 MAJ-188, MAJ-1614
WM-NP2, WM-DP2 MAJ-1639, MAJ-1650, MAJ-1653
Incorrect Correct