Chapter 3 Installation and Connection
47
HIGH FLOW INSUFFLATION UNIT UHI-4
Connecting the ESG-400 or the USG-400
1. Prepare the communication cable (MAJ-1872 or MAJ-1942 or MAJ-1948).
2. Connect the plug of the communication cable to the LINK-IN or LINK-OUT
terminal on the rear panel of the ESG-400/USG-400. After connection,
secure each plug by pushing it in while turning the screws on both sides of
the plug (see Figure 3.10).
3. Connect the plug of the communication cable to the LINK-IN or LINK-OUT
terminal on the rear panel of the high flow insufflation unit. After connection,
secure each plug by pushing it in while turning the screws on both sides of
the plug (see Figure 3.10).
Figure 3.10
Ultrasonic generator
(USG-400)
Electrosurgical generator
(ESG-400)
Incorrect connection
(This device and ultrasonic generator or electrosurgical
generator are connected with two communication cables.)
Correct connection
LINK-IN terminal
LINK-OUT terminal
LINK-IN terminal
LINK-OUT terminal
LINK-IN terminal
LINK-OUT terminal
LINK-IN terminal
LINK-OUT terminal
High flow
insufflation unit
(UHI-4)
High flow
insufflation unit
(UHI-4)
Ultrasonic generator
(USG-400)
Electrosurgical generator
(ESG-400)
Communication cable
Communication cable
Communication cable