Chapter 3 Installation and Connection
49
HIGH FLOW INSUFFLATION UNIT UHI-4
Figure 3.11
Convertor adapter for
UHI-2/3 (MAJ-1873)
System terminal
Gender changer or
Extension cable for
aeration (MAJ-1424)
Electrosurgical unit
(UES-40)
SonoSurg generator
(SonoSurg-G2)
Convertor adapter for
UHI-2/3 (MAJ-1873)
System terminal
Gender changer
Incorrect connection
(This device and electrosurgical unit or SonoSurg generator
are connected with two communication cables.)
Correct connection
LINK-IN terminal
LINK-OUT terminal
LINK-IN terminal
LINK-OUT terminal
High flow
insufflation unit
(UHI-4)
High flow
insufflation unit
(UHI-4)
Communication cable
Communication cable
Communication cable
Electrosurgical unit
(UES-40)
SonoSurg generator
(SonoSurg-G2)
Extension cable for
aeration (MAJ-1424)
or