108
Section 5 Linear Output
ZX-E
Operation Manual
Section 5
DETAILED SETTINGS
6. Press the ENT Key.
7. Use the LEFT and RIGHT Keys to display
L-ADJ on the main display.
The units for the monitor focus settings (mA or V) will be
shown on the sub-display.
8. Press the ENT Key.
The display will change to the settings for the first point
(A). The output current value will be displayed on the
main display and the correction will be displayed on the
sub-display and the leftmost digit will flash.
Setting the First Point (A)
9. Set the output current and correction values
for the first point.
Adjust the correction value on the sub-display so that the
ammeter reading and the output current shown on the
main display are the same. The larger the correction
value, the larger the output current.
The correction value can be set within the range
999 to
999. To set a negative value, make the leftmost digit of
the sub-display flash and change the value.
The flashing digit, i.e., the digit for which a value
can be set, will change as shown in the diagram.
10. Press the ENT Key to confirm the settings.
The correction value for the first point will be confirmed.
The screen for setting the second point correction value
will be displayed.
Move between digits.
Increment and decre-
ment the numeric value.
Sub-display
digit movement
Digit moves
one at a
time
Movement between main display
and sub-display (from leftmost or
rightmost digit of sub-display)
Z166-E1-02A.book Page 108 Friday, June 3, 2005 4:18 PM