122
Section 6 Measuring with Multiple Amplifier Units
ZX-E
Operation Manual
Section 6
AUXILIARY FUNCTIONS
Selecting the Number of Units Installed
10. Use the LEFT and RIGHT Keys to display
S-CH on the main display.
11. Press the UP or DOWN Key.
The sub-display will flash.
12. Use the UP and DOWN Keys to select the
total number of Amplifier Units.
Enter the value for the total number of con-
nected Amplifier Units. Mutual interference pre-
vention cannot be set for only some of the Units.
13. Press the ENT Key to confirm the setting.
The number of connected Amplifier Units will be regis-
tered.
Setting the Number of Samples to Average
14. Set the same number of samples to aver-
age for all Amplifier Units.
Setting Number of Samples to Average,
p. 80