92
Section 5 Changing Display Scales
ZX-E
Operation Manual
Section 5
DETAILED SETTINGS
One-point Scaling
With one-point scaling, measurement is performed for one position and an offset value
is set for that measurement. Both offset and display inversions (changing the relation-
ship between increasing and decreasing values) can be set.
This section describes how to make settings for one-point scaling, using an different
example.
Example: Displaying the Height of the Sensing Object
Moving to FUN mode and SPCL
1. Set the mode switch to FUN.
2. Use the LEFT and RIGHT Keys to display
SPCL on the main display.
Moving to SCALE
3. Press the UP or DOWN Key.
The sub-display will flash.
4. Use the UP and DOWN Keys to display
SET or ALL.
5. Press the ENT Key.
Sensor Head
Actual distance: 1.0 mm
Value to display: 1.5 mm
Display value (mm)
Actual measured
value (mm)
Sensing
object
Z166-E1-02A.book Page 92 Friday, June 3, 2005 4:18 PM