USER'S MANUAL
General instructions for use
(Rev. 0) I-Max Touch (110-120V)
99
6.10.1 Error message with error code E000 ÷
÷÷
÷ E199
NOT resettable errors.
These are internal errors of the control system; it is necessary to call the
Technical Assistance Service.
6.10.2 Error message with error code E200 ÷
÷÷
÷ E299
This category of errors apply to the rotation motor; of these only the error
"E206 - Collision with patient", caused by a possible collision between
the rotation arm and the patient, it is an actual reversible.
Press the key "Patient Entrance" (6) to reset the error and to
perform the axes centring operation.
For all other cases, call the Technical Assistance Service.