USER'S MANUAL
General instructions for use
I-Max Touch (110-120V) (Rev. 0)
74
6.7.1.2 Manual exposure
If the kV and mA combinations of the Table 4 are not considered suitable
for a specific examination, it will be possible to set new parameters using
the manual mode.
To modify the kV or mA values, press any of the up (3) or down
(4) arrows of the KV or mA parameters, the blue frames around the
"Examination mode selection", the "Adult/Child Selection" (7) and the
"Size Selection" (8) keys will disappear, orange frames will appear around
the up and down arrow keys of the parameters (3) and (4) and the
"Anatomic/Manual mode" indicator will display "M". The display will
show respectively one of the following two indications:
> x x k V x x m A 9 . 4 0 s
S I N U S
or
x x k V > x x m A 9 . 4 0 s
S I N U S
The symbol ">" indicates which parameter is being changed.
A parameter can be modified by pressing the increase key (3)
and the decrease key (4) of that parameter repeatedly.
The kV value can vary between 60 and 86 kV, with 2 kV steps.
The value of mA can vary between 6 and 10 mA, with 1 mA steps.
NOTE:
To change the values rapidly, keep the increase key (3) or decrease key
(4) pressed.
*