USER'S MANUAL
General instructions for use
(Rev. 0) I-Max Touch (110-120V)
85
6.8.2.1 Anatomic exposure
Select the type of patient with the Adult/Child key (7).
Select the type of build with the Size (8) key (small - medium - large).
The kV and mA values will be displayed according to the selections made
as per the following tables:
Adult Child
kV mA kV mA
Small 66 6 62 6
Medium 70 6 64 6
Large 72 6 66 6
Table 5: Latero-Lateral projection
Adult Child
kV mA kV mA
Small 74 12 70 10
Medium 76 12 72 10
Large 80 10 74 10
Table 6: Antero-Posterior projection