6.11.2 Defects due to wrong data setting
Problem Description Solution
Light or poorly
contrasted image.
Over-dark image.
The set kV values are not
adequate for the size of the
patient.
Try to modify the contrast, using the
appropriate commands of the image
acquisition/management program; if
necessary, repeat the examination,
varying the kV and/or the mA.
Increase them if the image was too
clear, and reduce them if it was too
dark.
If the error is repeated, call the
Technical Service.
Image completely
white.
No X-ray emission. Verify the emission of the X-rays by
acoustical and luminous signal. If
no solution can be found, call the
Technical Service.
The STF value is not correct. Refer to paragraph 6.8.3 to adjust
the position of the "STF".
Soft Tissue not (or
poorly) visible in L-L
projection.
A symmetric image format
was selected.
Select an asymmetrical image
format (which will enable the STF
filter).