USER'S MANUAL
General instructions for use
I-Max Touch (110-120V) (Rev. 0)
40
6.3 Switching on the device
Press the green button on the base of the column to switch the system
on; the display shows:
I M A X T O U C H
H W = x . x x S W = x x . x x
This message will be present for about 20 seconds.
After this time the LEDs on the control panel start blinking and on the
display will be present the following message:
I M A X T O U C H
R E L E A S E * . * *
3 seconds later, the display shows the following message:
> T E S T <
NOTE:
During this phase, the I-Max Touch does not perform any movement, it
just performs a series of checks which, in the event of negative result,
could require the intervention of the technician.
The only problem that can be solved by the user is related to the position
of the PAN sensor holder; in this case, the following message will be
displayed:
C L O S E C A S S E T T E
P A N O R A M I C
When the self-diagnosis is completed, the following appears on the
display:
M A C H I N E S E T T I N G
P R E S S > 0 <
Press key (6) to start the device alignment phase.
Once the key has been pressed, the message disappears and the display
shows the following message during the alignment of the axes:
W A I T F O R . . .
M A C H I N E S E T T I N G
WARNING:
During this phase, the machine checks for possible obstacles that may
create collisions simulating the movements performed during the
examination.
*