USER'S MANUAL
General instructions for use
I-Max Touch (110-120V) (Rev. 0)
102
6.10.4 Error message with error code E400 ÷
÷÷
÷ E402
NOT resettable errors.
These errors are related to the Soft Tissue Filter of the Digital CEPH.
Switch off the system and on again, in case of further error message, call
the Technical Assistance Service.
6.10.5 Error message with error code E700 ÷
÷÷
÷ E799
WARNING:
These error codes refer to the X-rays generation, therefore, they can also
indicate a safety problem.
With error code E759, turn off immediately the system as a not
requested X-ray emission was detected. In this case, call
immediately the Technical Assistance Service.
6.10.5.1 E755 – Safety Buckup Timer intervention
This message is signalling that the RX emission has not ended at the
correct time, but it has been terminated by the Safety Backup Timer.
This hardware device has interrupted the emission, but in any case
power off the system.
6.10.5.2 E774 - X-rays button not pressed
The lack of the button is signalled also if the emission software control is
present.
The error signals a possible failure on the connection of the X-rays
button with the generator card.
6.10.5.3 E775 - X-rays button released prematurely
The release of the X-rays button during the emission phase is signalled;
this signalling has a different meaning from that of the corresponding
E362 error, as this message is generated by the HF board, which signals
a possible failure on the connection of the X-rays button with the board
itself.