6.10.3 Error message with error code E300 ÷
÷÷
÷ E399
6.10.3.1 Error message with code error E300 ÷
÷÷
÷ E303
NOT resettable errors.
These errors are related to the secondary collimator of the Digital CEPH.
Switch off the system and on again, in case of further error message, call
the Technical Assistance Service.
6.10.3.2 Error message with code error E320 ÷
÷÷
÷ E323
NOT resettable errors.
These errors are related to the primary collimator.
Switch off the system and on again, in case of further error message, call
the Technical Assistance Service.
6.10.3.3 E340 - Sensor holder not in PAN position
A Panoramic type examination was requested, but the sensor holder does
not appear to be closed; close it in the PAN position and press the
key "Patient Entrance" (6) to reset the error condition.
6.10.3.4 E360 / E361 - X-ray button pressed during start up or axis
movement
Release the X-ray button if pressed; press the key "Patient Entrance"
(6) to reset the error condition.
If the error does not disappear, call the Technical Assistance Service.