6.13.1 Table of pre-set anatomic parameters
PANORAMIC TMJ open/close mouth
Adult
Child
Adult
Child
Small
68
6
kV
mA
64
6
kV
mA
Small
68
6
kV
mA
62
6
kV
mA
Medium
72
6
kV
mA
66
6
kV
mA
Medium
72
6
kV
mA
64
6
kV
mA
Large
74
6
kV
mA
68
6
kV
mA
Large
76
6
kV
mA
66
6
kV
mA
SINUS
Adult
Child
Small
66
6
kV
mA
62
6
kV
mA
Medium
70
6
kV
mA
64
6
kV
mA
Large
72
6
kV
mA
66
6
kV
mA
CEPHALOMETRY (L.L.) CEPHALOMETRY (A/P - P/A)
Adult
Child
Adult
Child
Small
66
6
kV
mA
62
6
kV
mA
Small
74
12
kV
mA
70
10
kV
mA
Medium
70
6
kV
mA
64
6
kV
mA
Medium
76
12
kV
mA
72
10
kV
mA
Large
72
6
kV
mA
66
6
kV
mA
Large
80
10
kV
mA
74
10
kV
mA