5. TECHNICAL CHARACTERISTICS
General features
Type I-Max Touch
Manufacturer OWANDY
Champs sur Marne, France
Class Class II B for European Directive
for Medical Devices 93/42
Class II for Canadian MDR
Class I with type B applied parts
according to IEC 60601-1
Class II according to 21CFR-
subchapter J
Protection degree IPX0 standard device
Rated line voltage
110-120 V
Line frequency 50/60 Hz
Maximum line current
15 A @ 115 V 50/60 Hz
Power consumption
1.7 kVA @ 115 V 50/60 Hz
Protection fuse (F1) 15 A T
Protection fuse (F2) of switching power
supply
3 A T
HF generator board protection fuses F1: 10 A F
F2: 5 A HF
F3: 2 A T
Line voltage regulation
< 3 % at 99 V
Rated output voltage (kVp) 60 ÷ 86 kV
p,
with 2 kV
p
steps
Anodic current 6 ÷ 10 mA, with 1 mA steps for PAN, TMJ,
and Sinus
6 ÷ 12 mA in 1 mA steps for Ceph (up to
76 kVp)
6 ÷ 10 mA in 1 mA steps for Ceph (from
78 kVp to 86 kVp)
Sensor cover additional filtration 0.5 mm Al eq @ 70 kV
p