USER'S MANUAL
General instructions for use
(Rev. 0) I-Max Touch (110-120V)
103
6.10.6 Error message with error code E850 ÷
÷÷
÷ E852
These errors signal abnormal situations due to the operator's interface.
6.10.6.1 E850 - One or more keys appear to be pressed on start-up
The system checks that all keys are not pressed at start-up; if one or
more appear to be pressed, this error is displayed.
If error E850 is detected, the display will show which key has been
pressed in start-up phase, and the following message will be shown:
E 8 5 0 ( x x x x x x x x )
S W I T C H P O W E R O F F
xxxxxxxx: error message code number
Release the key and switch ON again the unit. If the problem is still
present, call Technical Service.
6.10.6.2 E851 - Column key pressed
This error is displayed in case, when releasing the up/down column key,
the movement itself is not completed; pressing any other key interrupts
the movement to avoid injuries to the patient.
Press the key "Patient Entrance" (6) to reset the error condition.
6.10.6.3 E852 - Key "Patient Entrance" pressed during the movement
During the system movement, the keyboard is disabled, but
if the key "Patient Entrance" (6) is pressed the movement is
interrupted. This operation is useful in case a movement anomaly is
noticed.
Press the key "Patient Entrance" (6) to reset the error condition.