USER'S MANUAL
Safety information
I-Max Touch (110-120V) (Rev. 0)
6
WARNING: PRECAUTIONS WHILE USING LASER CENTRING
DEVICES:
• Always keep the room well lit.
• Do not look into the output windows of laser centring units.
• Do not stare at the reflections of laser pointers.
• Instruct the patient to keep his/her eyes closed as long as the laser
pointers are active.
• Before starting an examination, the patient must remove earrings,
glasses, necklaces and whatever else could reflect the laser beam or
be impressed on the radiographic image.
• Do not clean the openings of laser centring devices with tools that
could modify the optics. Any cleaning must be performed only by
authorised technicians
.
Operations other than those indicated could cause the ejection of
dangerous non-ionising radiation.
WARNING:
The USB port on the keyboard MUST NOT
be used with an external Hard
Disk with own mains connection. It has to be used only with USB Pen
Drives.
NOTE:
Patient environment is defined as a distance of at least 1.5 m (4.9 ft) from
the patient.
If the PC is positioned outside the patient environment when in use, the
PC has to be compliant to IEC 60950, otherwise the PC must complies
with medical standard IEC 60601-1.
NOTE:
When the unit is switched on, do not move the rotating arm or the
tube-head.
*
*