USER'S MANUAL
General instructions for use
(Rev. 0) I-Max Touch (110-120V)
111
Anterior teeth are enlarged and blurred
Figure 21 shows the result of this error.
Figure 21
Possible causes:
The patient it positioned too much backward.
Solution:
Check the patient's positioning by using luminous beams.