EasyManua.ls Logo

Owandy I-Max Touch - Page 32

Owandy I-Max Touch
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USER'S MANUAL
Technical characteristics
(Rev. 0) I-Max Touch (110-120V)
27
3. The KV and mA parameters can be modified by pressing the increase
key and the decrease key of the KV and mA on the display.
The parameters can vary within the limits shown in the following
table:
Parameter Minimum value Maximum value
kV 60 86
mA 6 12
Table 1
4. Perform an exposure; technical factors can be read on the measuring
instrument.
NOTE:
The performance is guaranteed only if the measurement of kV and time is
done with the invasive method, due to the fact that the non-invasive
method may introduce errors for instruments tolerance or wrong
measurement condition.
5. To end the control program, press any key other than the increase
and decrease keys ; the display will indicate:
x x k V x x m A x x . x s
P A N O R A M I C - S T D
and the unit will return to standard mode.
*

Table of Contents

Related product manuals